Chapter 2  Operating Your Frequency Counter
Using the Menu
2-42
Operating Guide
2
Some of the menu items listed on the previous page provide information 
only (no settings are required [or possible] for these), such as Battery 
Voltage, Operation Hours, and information that identifies the Counter 
(Agilent model number, firmware version number, serial number, and 
installed option codes). These menu options are described in the 
remainder of this chapter and also in “The Front Panel Menus at a 
Glance
” on page 1-4.
NOTE
Always terminate each setting you change in a menu option by pressing 
the 
Enter
 key. You can abort any change while the Menu is displayed by 
pressing the 
Reset/Local
 key or the 
Clear
 key. Both keys nullify the change 
you made to the current menu option, but they do not affect any of the 
other menu options. The 
Clear
 key terminates the current menu session, 
but the 
Reset/Local
 key does not.
Reference Oscillator (REF OSC)
By default, the Counter uses its internal 10 MHz reference oscillator 
(or the optional Oven Timebase, Option 001) as a timebase for all 
measurements, unless REF OSC is set to EXT, and it detects a 1, 2, 5, or 
10 MHz reference signal on the Reference connector on the back panel. 
If an external reference signal is present, you can force the Counter to use 
the internal reference oscillator by setting the REF OSC menu option 
to INT.
The available settings are EXT (external) and INT (internal), and the 
default setting is INT.
NOTE
When REF OSC is set to internal (INT), the Counter outputs a 10 MHz, 
1 V p-p signal on the Reference connector on the back panel.