Welch Allyn Medical Diagnostic Equipment 48700 用户手册

下载
页码 8
Halogen Exam Light III with Fiberoptic Light Pipe
Model #’s - 48700, 48722, 48724, 48726, 48728
Halogen Exam Light III with Fiberoptic Light Cable
Model #’s - 48780, 48782, 48784, 48786, 48788
Halogen Exam Light III with Neonate Light Pipe
Model #’s - 48760, 48762, 48764, 48766, 48768
Halogen Exam Light III Light Box with Power Cord
Model #’s - 48740, 48742, 48744, 48746, 48748
INTRODUCTION
Thank you for purchasing the Welch Allyn Halogen Exam Light III.  We believe 
that it is the highest quality product of its kind available in the world today.  
By following the simple guidelines found within this manual, you should be 
rewarded with years of dependable, trouble free performance.  If you have 
any questions or concerns please feel free to contact our customer or 
technical service departments at (800) 535-6663 during normal business 
hours. 
Welch Allyn remains committed to providing innovative, quality diagnostic 
products for healthcare professionals.  Call us at (800) 535-6663 to request a 
full line product catalog or visit us at our Web Site, www.WelchAllyn.com. 
SYMBOL DESCRIPTIONS
Attention: Consult user’s manual for additional information.
Warning: Risk of fire. Replace fuses as marked.
Caution: High Temperatures
Caution: High Intensity Light
WARNINGS & CAUTIONS
WARNING: The user of this equipment should be thoroughly trained 
in the medical procedures appropriate to the utilization of this 
instrumentation. Furthermore, time should be taken to read and 
understand the instructions contained within this manual prior to use 
of this product. Instructions for other equipment used in conjunction 
with the Exam Light III (for example: suction machines, 
electrosurgical generators, etc.) should also be read and understood. 
Failure to understand the operating requirements for this product 
may result in injury to yourself, the patient and/or may damage the 
instrument. 
CAUTION: Electric Shock Hazard - Remove power cord from 
electrical outlet and allow lamp to cool before replacing (Welch 
Allyn lamp # 04200).
DANGER: Risk of explosion if used in the presence of flammable 
anesthetics.
CAUTION: Federal Law restricts this device to sale by or on the order 
of a physician.
CAUTION: Illumination is intense - Do not stare at the light source in 
operation. May be harmful to eyes.
CAUTION: Use the examination light in its intended working range of 
12 to 24 inches. Exposures at closer distances may be harmful to 
skin.
WARNING: Do not perform neonatal trans-illumination without the 
use of the #48210 neonatal trans-illuminator (with U.V. filter). Do not 
remove the focusing sleeve from the light pipe and use the light pipe 
for trans-illumination. Severe burns may result.
General practices to minimize risk of harm to the skin from optical 
radiation hazards include: minimizing illumination intensity at the tis-
sue examination site, minimizing exposure times, and taking addi-
tional precautions when skin sensitivity has been altered through 
tissue trauma or the use of anesthesia.
General good practices to minimize risk of harm to the eyes from 
optical radiation hazards include: avoid looking at bright light sources 
and their reflections, and protect eyes where normal pupil sizes and 
aversion responses are not present.
There are no user serviceable/replaceable parts other than lamp and 
fuses. Please refer any other service to qualified and authorized 
service centers.
COMPONENTS
Standard
Light Box, Power Cord and Wall Mount
Model 48740 (Domestic)
Model 48742 (Europe)
Model 48744 (UK)
Model 48746 (Australia)
Model 48748 (Japan)
Fiberoptic Light Pipe (shown with focusing sleeve)
Model 48200 
(Included only with Model #’s 48700, 48722, 
48724, 48726, and 48728)
Fiberoptic Light Cable (shown with focusing sleeve)
Model 48220 
(Included only with Model #’s 48780, 48782, 
48784, 48786, and 48788)
Neonatal Transilluminator
Model 48210
(Included only with Model #’s 48760, 48762, 
48764, 48766, and 48768)
Focusing Sleeve
Model 48605
(Included only with Model #’s 48700, 48722, 48724, 
48726, 48728, 48780, 48782, 48784, 48786, and 48788)
Accessories
Fiberoptic Light Pipe (Shown with Focusing Sleeve)
Model 48200
Fiberoptic Light Cable (Shown with Focusing Sleeve)
Model 48220
Focusing Sleeve
Model 48605
Neonatal Transilluminator
Model 48210
The optional neonatal Light Pipe/
transilluminator is used to assist in the 
diagnosis of pneumothorax, to locate blood 
vessels, etc. This transilluminator 
incorporates a permanently attached 
ultraviolet filter that minimizes the 
transmission of ultraviolet light. The standard 
Fiberoptic Light Pipe or Fiberoptic Light Cable 
should NOT be used for transillumination of 
neonates. 
Table Mount Bracket
Model 48859
Mobile Stand
Model 48850
Disposable Sheaths
Model 52640
Disposable Sheath Dispenser
Model 52641
35W Halogen Replacement Lamp
Model 04200
MOUNTING 
Use one of the following three mounting options in order to ensure proper 
operation of the light box.
Wall Mount
Choose appropriate mounting location. The length of the light pipe is 4 feet 
totally extended (fiberoptic light cable is 7.5 feet). When choosing a mounting 
location, keep these dimensions in mind in addition to accounting for the 
actual application the light will be used for and what portion of the 
examination table will need to be accessed.
1.
Using the mounting plate as a template, mark drilling holes as needed 
(depending on your wall type).
2.
Dry Wall, Paneling, Plywood or Plaster 
(1/8 inch (.3 cm) to 5/8 inch (1.6 cm) 
thick) - drill four (4) 3/8 inch (1cm) holes, 
insert hollow metal wall anchor 
assembly. 
3.
Concrete Block - drill four (4) 3/16 inch 
(.5 cm) holes, insert plastic anchors. Use #8 x 1/2 inch screws to 
secure.
4.
Metal Panel or Plywood walls greater than 5/8 inch (1.60 cm) thick - 
drill four (4) 1/8 inch (.3 cm) holes. Use #8 x 1 inch screws to secure. 
5.
Once holes are drilled, install the mounting plate with the appropriate 
screws. Completely tighten the screw so there is no gap between the 
screw head and the mounting plate.
6.
Mount the Light Box to the plate by inserting the pins on the back of the 
light box into the slots on the mounting plate. Let the light box drop 
down to position the pins into the narrow portion of the mounting slots 

. 
7.
To secure the Light Box to the plate, insert two (2) .50” (1.27cm) screws 
through the bottom of the mounting plate and tighten into the bottom 
of the Light Box using a phillips head screwdriver 

.
NOTE:  Some exam table manufacturers supply mounting hardware for 
installation of the Exam Light III. Check with your table supplier for 
availability.
WARNING:  It is the responsibility of the health facility of user to ensure that 
the mounting bracket  is securely anchored to the wall or table, or that the 
mobile stand is securely assembled, and the exam light is securely fastened 
to the bracket or stand as directed below. Give consideration to the load that 
is placed upon the mount.  Welch Allyn, Inc. does not accept responsibility for 
any installation, nor for damage or injury arising from the installation of any 
mounting bracket or mobile stand, regardless of the type of fasteners and 
technique used.
1.
Determine the side of the table that is most optimal for mounting the 
light. Consider the location of the nearest power outlet and patient 
traffic.
NOTE:  Verify that the top of the mounting bracket is at least 10 inches (25.4 
cm) below the table top cushion and that the mounting position will not 
interfere with the operation of the table.
2.
Using the mounting plate as a template, mark the hole locations on the 
table and drill four (4) .25 inch (.64 cm) holes in the table.
3.
Using four 8-32 screws with lock washers and nuts, install the 
mounting plate to the table as shown in the diagram.
4.
Mount the light box to the plate by inserting the pins on the back of the 
light box into the slots on the mounting plate. Let the light box drop 
down to position the pins into the narrow portion of the mounting slots.
5.
To secure the Light Box to the mounting plate, insert two (2) .50” (1.27 
cm) screws through the bottom of the mounting plate and tighten into 
the bottom of the Light Box using a Phillips head screwdriver.
NOTE: Some tables require additional reinforcement of the mounting bracket 
to support the mounted light. In this case, we recommend using a 1” (2.5 cm) 
thick piece of plywood. When employing this reinforcement to the bracket, 
please follow instructions 6 - 11 below.
6.
Trace the outline of the mounting plate and mark the hole locations on 
a 3/4 inch (1.9cm) thick piece of plywood.
7.
Cut the piece of wood to the outline size and drill four .25 inch (.64cm) 
holes.
8.
Using the mounting plate as a template, mark the hole locations on the 
table and drill (4) .25 inch (.64 cm) holes in the table. 
9.
Using four 8-32 x 2 inch (5.08 cm) screws with lock washers and nuts, 
install the block of wood and mounting plate to the table as shown in 
the diagram.
10.
Mount the light box to the plate by inserting the pins on the back of the 
light box into the slots on the mounting plate. Let the light box drop 
down to position the pins into the narrow portion of the mounting slots.
11.
To secure the Light Box to the mounting plate, insert two (2) .50” 
(1.27cm) screws through the bottom of the mounting plate and tighten 
into the bottom of the Light Box using a phillips head screwdriver.
Halogen Exam Light III