Sekonic L-398M 用户手册

下载
页码 36
 
 
 
 
 
 
2. Lumidisc.
  Employ for illumination contrast 
Adjustment and luminance measurement. 
 Contrast measurement 
a. Mount Lumidisc 
sensor 
Install in the same 
manner as Lumisphere. 
b. Hold meter at subject position 
and point Lumidisc 
directly toward center of main light. 
c. Press stopper button 
read meter scale footcandle 
indication 
d.  Turn the memo pointer dial and set the 
pointer 1 7 at the footcandle value the 
needle points to 
e.  In the same manner, measure the fill-in 
light. In this case, use care that light from 
main source dose not enter sensor. 
f. 
Read a difference between the 
footcandle value the pointer is set (see 
above c) and the one for fill-in light 
(= 
needle position. 
Ratio between main and fillip light 
footcandles becomes the contrast ratio. 
Example: 
With main light at 640 
footcandles, fill-in light reads 
320 footcandles. Illumination 
contrast ratio becomes: 
 
  640 
320 
1   or 2 : 1 
 
 
 
11