Technical Specifications
English
 Displayed values
 Maximum low impedance flow at stated pressures
Glossary of symbols
Read the relevant operating manual before using
Type CF equipment
AC switch
Class II
Drip proof
Value
Range
Accuracy
Display 
Resolution
Proximal cuff
Pressure
-3 to 40 cm H
2
O
0.5 cm H
2
O
0.1 cm H
2
O
Mass flow sensor
Leak
0–99 liters per minute
+/- 12 L/min
1 liter per breath
Respiratory rate
8–30 breaths per minute
+/- 10%
1 breath per minute
Tidal volume
0–4500 milliliters
+/- 10%
1 milliliter
Minute ventilation
0–40 liters per minute
+/- 10%
0.1 liters per minute
Target MV
0–40 liters per minute
+/- 10%
0.1 liters per minute
Pressure (cm H
2
O)
Flow (L/min)
4.2
217
8.6
218
13.0
219
IPX1