Philips HEARTSTART XLT M3500B 用户手册

下载
页码 180
Connecting to Power
2-9
2
Connecting to Power
The HeartStart XLT is powered by the M3516A battery. Prior to inserting the 
battery, make sure that the battery is charged and has been properly main-
tained (See “Battery Maintenance” on page 11-8). A fully charged battery will 
last for about two hours. A second spare charged battery should be kept in the 
carrying case at all times. 
The HeartStart XLT can also be powered by the M3517A AC Power Module 
or the M3518A DC Power Module. However the defibrillator will take longer 
to charge when powered by a power module, with the battery absent. The rec-
ommended practice when using a power module is to use the module in con-
junction with a battery.  For information on using the power modules, see the 
documentation supplied with the modules.
NOTE
To save battery capacity, the HeartStart XLT shuts itself off if a patient is not 
being monitored and no one has interacted with the device for 10 minutes.
Inserting the Battery
To insert the battery, slide it into the battery receptacle as shown in Figure 2-6. 
Then push the battery in until you hear an audible click.
Figure 2-6  Inserting the Battery