ERC Protocol
3-16
Defibrillating in AED Mode
ERC Protocol
The HeartStart XLT can be configured to support the European Resuscitation 
Council (ERC) Guidelines for Resuscitation (1998). If 
(XURSHDQ3URWRFRl is con-
figured to 
2Q, the defibrillation process described in this chapter is the same, 
with the exception of how the Pause state functions (see “Pausing for CPR” 
on 3-12). 
As described, you can enter a Pause state:
z
at the completion of a shock series, or
z
when no shock is advised
When either of these events occurs, the ERC protocol prompts you to Check 
Patient. Then it prompts you as follows:
Figure 3-14  
NOTE
Using the ERC protocol, you are not prompted to (or given time to) check the 
patient’s pulse.
Shocks: 1
00:00:50
If Needed
Press PAUSE
And Begin CPR
PAUSE
ANALYZE