Philips HEARTSTART XLT M3500B 用户手册
ERC Protocol
3-16
Defibrillating in AED Mode
ERC Protocol
The HeartStart XLT can be configured to support the European Resuscitation
Council (ERC) Guidelines for Resuscitation (1998). If
Council (ERC) Guidelines for Resuscitation (1998). If
(XURSHDQ3URWRFRl is con-
figured to
2Q, the defibrillation process described in this chapter is the same,
with the exception of how the Pause state functions (see “Pausing for CPR”
on 3-12).
on 3-12).
As described, you can enter a Pause state:
z
at the completion of a shock series, or
z
when no shock is advised
When either of these events occurs, the ERC protocol prompts you to Check
Patient. Then it prompts you as follows:
Patient. Then it prompts you as follows:
Figure 3-14
NOTE
Using the ERC protocol, you are not prompted to (or given time to) check the
patient’s pulse.
patient’s pulse.
Shocks: 1
00:00:50
If Needed
Press PAUSE
And Begin CPR
PAUSE
ANALYZE