Konica Minolta PULSOX 300 用户手册

下载
页码 32
24
Appendix
Probes (SR-5C, SP-5C, LM-5C, SD-5C)
Probes (SR-5C, SP-5C, LM-5C, SD-5C)
■ 
LED-SPD probe
■ 
Type
•  Finger Clip Probe SR-5C
•  Spot Check Probe SP-5C
•  Monitor Probe LM-5C
•  Monitor Probe LM-5C
•  Personal Probe SD-5C
✻ 
It can be used with one of Extension Cable EC-300
■ 
Components that touch the body
Cover: 
Polycarbonate
Wrist band: 
Polyethylene terephthalate, nylon 6
Rating nameplate:  Polyethylene terephthalate
Cover
Wristband
Rating nameplate
Equipment classifi cation (based on “UL 60601-1/EN 60601-1”)
•  Protection against electric shock: Internally powered
•  Type of applied part: BF
•  Not suitability for use in the presence of fl ammable anaesthetic mixture 
with air or oxygen or nitrous oxide.
•  The instrument is not protected against entry of water.
•  Only the cleaning method has been stipulated (no disinfection/sterilization 
methods have been stipulated).
•  Mode of operation of Equipment: Continuous while in Use (IEC 60601-1)