Analog Measurements - Screens and Control Fields
Normalize
Notch Freq
Normalize
This area performs three display operations: 
Save B
 saves the currently-displayed trace for the A-B operation. 
A only
 provides a continuously-updated display (the “normal” mode 
of operation). 
A-B
 displays the difference between the trace saved using 
Save B
 
and the currently displayed trace. The comparison can yield either 
losses or gains in amplitude.
Operating Considerations
The A-B function works correctly only if the center frequency and span 
settings are the same for both signals. 
The reference level (
Ref Level
) can be changed (on the 
Main
 controls 
menu) to move the trace below the top line of the display if the A-B 
function results in a gain. 
The 
Hold
 key can be used to “freeze” the display at any time. This allows 
you to view a trace before performing the Save B or A-B functions. 
Screens on which this field is displayed
SPEC ANL: auxiliary controls menu
Notch Freq
This field sets the center frequency for the variable frequency notch 
filter. It is typically used for distortion and SINAD measurements at 
frequencies below or above the standard 1 kHz notch filter. 
Operating Considerations
When the 
Notch Coupl
 field on the INSTRUMENT CONFIGURE 
screen is set to 
AFGen1
, this filter and the 
AFGen1 Freq
 field match 
their settings. A warning message is then displayed if you attempt to 
set the 
AFGen1 Freq
 value outside the 300 Hz to 10 kHz range of this 
filter. When the 
Notch Coupl
 field is set to 
None
, this filter and 
AFGen1 Freq
 operate independently. 
Screens on which this field is displayed
AF ANALYZER
Main Menu