Operating Overview
Calibrating the Channel Power Measurement
Calibrating the Channel Power Measurement
To calibrate a channel power measurement at the RF IN/OUT port
using the internal CDMA generator
using the internal CDMA generator
Step 1. Go to the CDMA ANALYZER screen and choose the
Chan Pwr
measurement.
Step 2. Set the
Tune Freq
field (or
RF Channel
field) to the center
frequency of your CDMA signal.
Step 3. Select the
Chn Pwr Cal
field and wait until the calibration is
complete.