We reserve the right to technical modifications without prior notice.
GE Inspection Technologies
Xray
©2004 General Electric Company. All rights reserved.
   GEIT-30134GB (6/04)
GEInspectionTechnologies.com
The generation of radiation in an X-ray 
tube solely depends on the operation 
values, not on the make. 
The dose rate relevant in practice and 
suitable for calculations of radiation pro-
tection values is defined by national stan-
dards; thus the dose rate of the tubehous-
ing MXR 320/23, measured at a distance 
of 1 m from the focal spot, amounts to 
10.17 Sv/h at maximum tube voltage and 
maximum anode dissipation.
This value must not be used to assess 
biological effects.
The dose rate of the leakage radiation is  
< 5 mSv/h (1 rem/h).
Dose Rate within the Central Beam
Technical Data
10
15
70
65
20
25
30
35
40
45
50
55
60
95
90
75
80
85
110
105
100
D7
Belichtungsdiagramm
FFA = 900 mm  D7Pb  S = 2,0
Exposure Diagram
FFD = 900 mm  D7Pb  D = 2.0
mA x min
mm Fe
450 k
V
420 k
V
400 k
V
380 k
V
360 k
V
340 k
V
320 k
V
300 k
V
280 k
V
26
0 k
V
240 k
V
220 k
V
200
20
50
10
5
1
2
100
Maximum tube voltage 
320 kV
 
Large focal spot 
Small focal spot
Maximum anode dissipation 
1600 W 
640 W
Tube current at max. tube voltage 
5 mA 
2 mA
Focal spot size (EN 12 543) 
3.60 mm (~ 1.8 IEC 336) 
1.90 mm (~ 0.8 IEC 336)
Emergent beam angle 
40°
Inherent filtration 
3 mm Be
High voltage connection 
2 Plug sockets for rubber cone plugs R24
 
with optional quick-lock cable flanges
Cooling oil flow rate 
min. 14 l/min
Cooling oil temperature 
max. 50° C
Cooling oil pressure 
max. 7 bar
Weight (with optional cable quick-lock) 
40 kg (88 lbs)
Dimensions 
see drawing
0 150 mm
   (5.9")
96 mm  (3.8")
385 mm (15.2")
273 mm (10.7")
546 mm (21.5")
0
2
4
6
8
10
12
14
16
18
20
22
mA
0
50 100 150 200 250 300 350
kV