8. Troubleshooting
Check the following if your unit should fail during operation. You can also refer to 
the pages of this manual for complete instructions.
Notes:
• If the suggested remedy does not solve the problem, do not try to repair the 
device - no parts of the unit are user serviceable. 
Return the unit to an authorized OMRON retail outlet or distributor.
• The device should be serviced once a year by an approved OMRON service 
centre.
Symptom
Cause
Remedy
Nothing happens when the 
power switch is pressed.
Is the power plug plugged 
into an electrical outlet?
Check that the plug is 
inserted in an electrical 
outlet. Unplug then reinsert 
the plug if necessary.
No nebulisation or low 
nebulisation rate, when the 
power is on.
Is there medication in the 
medication tank?
Fill the medication tank.
Is there too much/little 
medication in the medication 
tank?
Add the correct amount of 
medication to the medication 
tank.
Is the vaporiser head 
missing and not assembled 
correctly?
Attach the vaporiser head 
correctly. 
Is the nebuliser kit 
assembled correctly?
Assemble the nebuliser kit 
correctly.
Is the nozzle blocked?
Make sure that the nozzle is 
free of blockages.
Is the nebuliser kit tilted at a 
sharp angle?
Make sure that the nebuliser 
kit is not tilted at an angle of 
more than 45 degrees.
Is the air tube connected 
correctly?
Make sure that the air tube is 
correctly connected to the 
compressor and nebuliser.
Is the air tube folded or 
damaged?
Make sure that the air tube 
does not contain kinks.
Is the air tube blocked?
Make sure that the air tube is 
free of blockages.
Is the air filter dirty?
Replace the air filter with a 
new one.
The device is abnormally 
loud.
Is the air filter cover attached 
correctly?
Attach the air filter cover 
correctly.
The device is very hot.
Are the ventilation slots 
blocked?
Do not block the ventilation 
slots.