NI PXI-1006 Power Supply Shuttle User Guide
4
ni.com
Connecting Safety Ground
Caution
The power supply shuttle is designed with a three-position NEMA 5-15 jack that 
connects the ground line to the chassis ground. To minimize shock hazard, make sure the 
electrical power outlet you use to power the chassis has an appropriate earth safety ground.
If your power outlet does not have an appropriate ground connection, you 
must connect the premise safety ground to the chassis ground screw. Refer 
to Figure 1 to locate the chassis ground screw. Complete the following 
steps to connect the safety ground.
1.
Connect a 16 AWG (1.3 mm) wire to the chassis ground screw using a 
grounding lug. The wire must have green insulation with a yellow 
stripe or must be noninsulated (bare).
2.
Attach the opposite end of the wire to permanent earth ground using 
toothed washers or a toothed lug.
Specifications 
AC Input
Input voltage range .................................90 to 264 VAC
Input frequency range .............................47 to 63 Hz
Maximum steady state
operating current.....................................10 A
Over-current protection ..........................10 A circuit breaker
Line regulation........................................±0.1% over operating line range
Efficiency................................................70–80% typical
Power disconnect ....................................The Off (standby) power switch 
causes the power module to 
supply DC power to 
the CompactPCI/PXI backplane. 
The rear-panel D-sub connector 
facilitates remote inhibiting 
operation. The Off (standby) 
switch must be in the On position 
prior to use of remote inhibit. The 
power cord provides main power 
disconnect.