Specifications 
Number of Turns 
50 
Maximum Current 
11A rms, continuous 
20.5A rms, 2 minutes 
Maximum Duty Cycle Derating 
< 11A, continuous 
> 11A, 2 minutes ON, 8 minutes OFF 
Maximum Voltage 
7 V dc or 7 V rms ac 
Minimum Inner Diameter of Clamp Jaws 
2.54 cm [1”] 
Specification for Clamp Meter/Coil interaction 
Specification for 5500A/COIL 
Coil Specifications [clamp/coil interaction] 
Calibrator Output 
Effective 
Current Output 
Toroidal-type Clamps 
[such as the Fluke 80i-600 and 80I-1000]
 
'other-type' of Clamps 
[such as the Fluke 80i-KW, -400, 
80i-410, -500, -1010 and 30-series]
 
Magnitude 
Frequency 
Amp-turns 
±(% of output + Amps) 
±(% of output + Amps) 
0.2 - 0.33 A 
DC 
10 - 16.4999 
0.50 
0.02 
0.33 - 2.9999 A 
DC 
16.5 - 149.999 
0.5 
0.14 
3.0 - 20.5 A 
DC 
150 - 1025 
Not applicable 
0.5 0.5 
0.2 - 0.33 A 
45 Hz to 65 Hz 
10 - 16.4999 
0.28 
0.003 
0.56 
0.03 
0.33 - 2.9999 A 
45 Hz to 65 Hz 
16.5 - 149.999 
0.28 
0.025 0.56 0.25 
3.0 - 20.5 A 
45 Hz to 65 Hz 
150 - 1025 
0.28 
0.09 
0.56 
0.9 
0.2 - 0.33 A 
65 Hz to 440 Hz 
10 - 16.4999 
0.79 
0.003 
1.00 
0.03 
0.33 - 2.9999 A 
65 Hz to 440 Hz 
16.5 - 149.999 
0.79 
0.027 1.00 0.25 
3.0 - 20.5
** 
65 Hz to 440 Hz 
150 - 1025 
0.79 0.1 
1.00 
0.9 
0.2 - 20.5 A 
Above 440 Hz 
10 - 1025 
Usable but unspecified 
**Note: Some calibrators may not have adequate compliance voltage to drive coil in this range. 
Calculating Total Specification 
The total specification of the effective current that the clamp-type meter measures is a function of the 
clamp/coil interaction and of the current calibrator. To determine the total specification use the following 
formula: 
 
source
2
coil
2
total
U
S
S
+
=
 
Example: 
Assume we are driving the coil with the Fluke 5500A at 4A, 60 Hz (the clamp meter will see an effective 
current of 200A, 60 Hz) and we are calibrating a toroidal-type clamp meter.  The Calibrator’s 1 year 
specification at 4 Amps is 
± (0.06% + 2 mA), so the effective current in the coil bundle will have a speci-
fication of 
± (0.06% + 0.1A).  Next, we find the total specification of the calibrator and the coil as a per-
centage of the output: 
Specification of effective calibrator current in coil bundle = 
± (0.06% + 0.1A) = 0.11% 
Specification due to clamp Meter/Coil Interaction = 
± (0.28% + 0.09A) = 0.325% 
 
The RSS of these two specifications determines the total specification of the clamp/source combination: 
%
343
.
0
%
11
.
0
%
325
.
0
2
2
total
=
+
=
S