10. In the Properties window, make the following se-
lections:
Bits per second
19200 (See Note)
Data bits
8
Parity
None
Stop bits
1
Flow control
Xon / Xoff
11. After making the selections, click on the
OK
but-
ton.
12. Press <ENTER> on the keyboard.
13. Verify that the
$
prompt appears on the PC dis-
play.
14. This completes the initial setup for calibration.
4-12
682XXB/683XXB MM
CALIBRATION
INITIAL SETUP
NOTE
When calibrating instruments with
operating firmware below Version
2.00, set the Bits per second rate to
9600. (This does not apply to
682X5B/683X5B models.)