General Troubleshooting & Repair of the M400E Analyzer 
M400E Ozone Analyzer Operator’s Manual 
13.7.7. MOTHERBOARD  
13.7.7.1. Test Channel / Analog Outputs Voltage 
The 
ANALOG OUTPUT submenu, located under the SETUP  MORE  DIAG menu is used to verify that the 
M400E analyzer’s three analog outputs are working properly.  The test generates a signal on all three outputs 
simultaneously as shown in the following table: 
For each of the steps the output should be within 1% of the nominal value listed in the table below except for the 
0% step, which should be within 0mV ±2 to 3 mV.  Make sure you take into account any offset that may have 
been programmed into channel (See Section 7.4.5).   
Table 13-9:  Analog Output Test Function - Nominal Values Voltage Outputs 
 
 
FULL SCALE OUTPUT OF  VOLTAGE RANGE 
(see Section 7.4.3
   100MV 1V 
5V 
10V 
STEP 
NOMINAL OUTPUT VOLTAGE 
1 0 
0  0  0 
2 20  20 
mV  0.2  1  2 
3 40  40 
mV  0.4  2  4 
4 60  60 
mV  0.6  3  6 
5 80  80 
mV  0.8  4  8 
6 100  100 
mV  1.0  5  10 
 
If one or more of the steps fails to be within these ranges, it is likely that there has been a failure of the either or 
both of the DACs and their associated circuitry on the motherboard.  To perform the test connect a voltmeter to 
the output in question and perform an analog output step test as follows:   
 
250 
04315 Rev. C1