Teledyne 400E 用户手册
General Troubleshooting & Repair of the M400E Analyzer
M400E Ozone Analyzer Operator’s Manual
13.7.7. MOTHERBOARD
13.7.7.1. Test Channel / Analog Outputs Voltage
The
ANALOG OUTPUT submenu, located under the SETUP MORE DIAG menu is used to verify that the
M400E analyzer’s three analog outputs are working properly. The test generates a signal on all three outputs
simultaneously as shown in the following table:
simultaneously as shown in the following table:
For each of the steps the output should be within 1% of the nominal value listed in the table below except for the
0% step, which should be within 0mV ±2 to 3 mV. Make sure you take into account any offset that may have
been programmed into channel (See Section 7.4.5).
0% step, which should be within 0mV ±2 to 3 mV. Make sure you take into account any offset that may have
been programmed into channel (See Section 7.4.5).
Table 13-9: Analog Output Test Function - Nominal Values Voltage Outputs
FULL SCALE OUTPUT OF VOLTAGE RANGE
(see Section 7.4.3)
100MV 1V
5V
10V
STEP
%
NOMINAL OUTPUT VOLTAGE
1 0
0
0 0 0
2 20 20
mV 0.2 1 2
3 40 40
mV 0.4 2 4
4 60 60
mV 0.6 3 6
5 80 80
mV 0.8 4 8
6 100 100
mV 1.0 5 10
If one or more of the steps fails to be within these ranges, it is likely that there has been a failure of the either or
both of the DACs and their associated circuitry on the motherboard. To perform the test connect a voltmeter to
the output in question and perform an analog output step test as follows:
both of the DACs and their associated circuitry on the motherboard. To perform the test connect a voltmeter to
the output in question and perform an analog output step test as follows:
250
04315 Rev. C1