The amount of volume loss you observe and the change in reactant concentrations you can 
tolerate determine the minimum volume that can be used. Typical volume losses of 1µl in 30 
cycles allow the use of samples of 20µl or less. 
During the final cool-down, a ring of condensation may form above the liquid level but below the 
top of the sample block. This is normal and does not effect the reaction. 
0.5ml Tubes
 
The heated lid mechanism of the Cycler TCY with the 0.5ml block is optimised for flat top tubes. In 
order to fit the maximum number of tubes in the block, they need to be placed so that all the 
hinges are in the same relative orientation. An example is illustrated below.  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Place the tubes in the block in the same 
orientation in order to achieve the maximum 
capacity. 
 
Always make sure the pressure on the 
heated lid is balanced by spreading the tubes 
across the block or inserting some dummy 
tubes if necessary.