11. Some frequently asked questions
 
 
Q1 
What is the ‘Pause’ function at the start of the program used for?  
A1 
Some users prefer to preheat the heated lid before placing the samples into the 
unit.  The  pause  feature  is  used  to  stop  the  unit a fter  the  4 minut e  heated  lid 
preheat step.  It will also sound an audible alarm indicating that the machine is 
ready  for the   sample  tubes  or pla te  to  be  added.  Pressing  the  “Pause”  key 
followed by the “Enter” key will commence the remaining program.   
 
 
Q2 
Why do you have to press some keys twice to make the unit do what you want 
it to do? 
A2 
With  many  of  the  program f eatures  there  is  a  fail-safe  function  which  asks  if 
you are sure you want to perform a particular action. You are required to press 
one key to select a function followed by a second to accept it.  
 
 
Q3 
What is ‘Hot Start’? 
A3 
A  “hot  start”  is  used  for  reducing  non-specific  product for mation  during  the 
first cycle of amplification. The ‘Hot S tart’ programming step is used to pause 
the instrument at a specific temperature, typically around 70ºC, after the initial 
template  denaturation.  The  reason  is  to  allow  the  manual  addition of  
unmodified Taq DNA polymerase which may loose activity if added during the 
initial  5min  denaturation.  Heat-activated  Taq  or  Hot  Start enzy mes  do  not 
require this step. 
 
 
Q4 
What is the incremental function? (See page 24). 
A4 
Incremental timing and temperature are used to increase or decrease either the 
time or temperature incrementally over the number of cycles in a stage.  
 
Incrementation of extension time is used with ‘Long PCR’ which is when large 
template fragments are to be amplified (e.g. 27kb lambda DNA, 40kb ge nomic 
DNA).  
 
Decremental  temperature  is  used  for  protocols  such  as  ‘Touchdown  PCR’ 
where  one  starts  with  a  high  annealing  temperature  in  the  first  cycle  and 
gradually  decreases  the  temperature  over  the numbe r  of  cycles  in the   stage. 
This ensures that only the specific product is amplified. 
 
 
Q5 
The  heat-up  ramp-rate  in  the  specification  is  3.6°C/second  yet  I  can  only 
manually set it to 2.0°C/second. Why is this? 
A5 
If you set the ramp-rate to max the block will heat at 3.6°C/second the unit sets 
its own control parameters. If you program a ramp-rate the control parameters 
reduce the ramp-rate. This is a feature of the software and not the actual ramp-
rate.