Omega Engineering 20 用户手册
7
3. Cycler TCY Specification
Temperature
Temperature range
4°C to 99°C
Temperature set point precision
0.1°C
Block uniformity (over full range)
± 0.5°C
Temperature accuracy
± 0.5°C
End of program cooling below ambient
Yes
Heating/Cooling Rate
Maximum heating rate
4.5ºC/s
Minimum heating rate
0.1ºC/s
Maximum cooling rate
3.0ºC/s
Minimum cooling rate
0.1ºC/s
Heated Lid
Heated lid enable/disable
Yes
Selectable heated lid temperature
100°C to 115°C
Over-temperature cut-out
Fixed at 145°C
Temperature
sensor
Thermistor
Heater
type
Etched
foil
Heater
power
33W
Warm up time
4 min approximately
(ambient to 105°C)
The heated lid will only come on if the set temperature is above 35°C
Programming
Program naming using alpha keys
Yes
Program password protection
Yes
Number of programs (3 step)
37
Maximum number of steps per stage
35
Maximum number of stages per program
25
Minimum hold time
1 sec
Maximum hold time
18 hrs
Incremental/decremental temperature
Yes
Incremental/decremental step hold time
Yes
Programmable ramp rate (heat or cool)
Variable in 0.1°C/sec steps
Programmable ramp rate resolution
0.1°C
Running Programs
Pause button
Yes
Stop button
Yes
End of program alarm
Yes
Auto restart on power fail
Yes
Interconnectivity
Connection to PC control program
Yes
Software updates from web-site
Yes