HP (Hewlett-Packard) E1368A 用户手册

下载
页码 86
OUTPut
The 
OUTPut
 subsystem enables or disables the "Trig Out" port of the HP
E1300/E1301 Mainframe or on the HP Command Module. 
Subsystem Syntax
OUTPut
[:STATe] <
state
>
[:STATe]?
[:STATe]
OUTPut[:STATe] <
state
>
  enables/disables the "Trig Out" port on the rear
panel of the HP E1300A/E1301A Mainframe or on the HP Command
Module.  
OUTPut[:STATe] ON | 1 enables the port and OUTPut[:STATe]
OFF | 0
 disables the port.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
state
boolean
0 | 1 | ON | OFF
OFF | 0 
Comments
Enabling "Trig Out" Port: When enabled, a pulse is output from
the "Trig Out" port after each switchbox channel is closed.  If
disabled, a pulse is not output from the "Trig Out" port after channel
closures.  The output pulse is a +5V negative-going pulse.
"Trig Out" Port Shared by Switchboxes: When enabled, the "Trig
Out" port may be pulsed by any switchbox each time a scanned
channel is closed.  To disable the output for a specific switchbox,
send the 
OUTPut[:STATe] OFF
 or 
OUTPut[:STATe] 0
 command for
that switchbox.
Related Commands: 
[ROUTE:]SCAN
TRIGger:SOURce
*RST Condition: 
OUTPut[:STATe] OFF
 (port disabled)
Example
Enabling "Trig Out" Port
OUTP:STAT ON
Enables "Trig Out" port to output
pulse after each scanned channel is
closed.
[:STATe]?
OUTPut[:STATe]?
  queries the present state of the "Trig Out" port.  The
command returns 1 if the port is enabled, or 0 if disabled.
Example
Query "Trig Out" Port Enable State
This example enables the "Trig Out" port and queries the enable state. The
OUTPut[:STATe]?
 command returns 1 since the port is enabled.
OUTP:STAT ON
Enables "Trig Out" port for pulse
output.
OUTP:STAT?
Query port enable state.
Chapter 4
HP E1368A/69A/70A Command Reference   57