XES Synergix 8825/8830/8850/8855 Digital Solution User Guide
118
Cleaning and maintenance - Scanner (continued)
Cleaning the Platen Glass and
Document Drive Rolls
Perform the following steps daily:
1.  Open the Top Cover.
2.  Dampen a clean, lint-free towel (600S4372) with platen glass cleaning
fluid (Xerox 43P81).
3.  Wipe the Platen Glass with the towel.
4.  Visually inspect the Platen Glass to ensure that all dirt, glue, and other
residue have been removed.
5.  Place a small amount of the cleaning fluid on the towel and clean the
front and rear Document Drive Rolls. Rotate the Rolls as you clean them
to ensure that the entire roll surface is clean.
6.  Place a small amount of the cleaning fluid on the towel and wipe the
white Backer Roll.
NOTE:  The Backer Roll is located on the underside of the Top Cover.  The
Backer Roll may rotate as you clean it.  This is normal behavior.
7.  Close the Top Cover.
Cleaning the Calibration Strip
Perform the following steps monthly:
1.  Run the Calibration Motor component test with FUNCTION set to GO TO
CALIBRATE and MOTOR set to ON.  Running this test with these
settings will position the Calibration Strip so that you can access it for
cleaning.  See "Calibration Motor" in the "Main Menu" section for
information about accessing and running this test.
NOTE:  The Top Cover must be closed while running this test.
2.  Raise the Top Cover.
3.  Inspect the Calibration Strip.
4.  Place a small amount of platen glass cleaning fluid (Xerox 43P81) on a
clean, lint-free towel (600S4372) and clean the Calibration Strip.
5.  Lower the Top Cover.
6.  Press the Exit key to exit the CALIBRATION MOTOR screen.
7.  Press the Menu key to return to the READY screen.  The Calibration Strip
automatically returns to its home position.