In steps 1 to 3 only data for the HTL (High Threshold Level), LTL (Low Threshold 
Level), or 3 dB (No Thresh-old) will he displayed or printed. In step 4 data for HTL, 
LTL, and 3 dB will be printed. Step 5 dumps raw data to the Computer. Step 6 reads or 
resets the Clock. Step 7 reads or stores the Calibrator level. 
 
 
Sound Level Meter Keys 
SOUND              Sound Level in decibels. Will repeat each second until
 
LEVEL                another key is pressed. To conserve power, press RUN,
 
PAUSE or any other key to exit sound level when it is
 
not needed. 
 
PEAK                  The highest unweighted Peak Level in decibels during
 
LEVEL                Run Time or an Event. In 30 dB range lowest peak 
 
measurable is 78 dB; in 50 dB range the lowest measurable is 98 dB.
 
MAX                    The Maximum Level in decibels after the filter and Slow
 
LEVEL                or Fast time constant circuit. 
 
Dosimeter Keys 
L-AVG                The average integrated sound level in decibels for the 
LEQ 
HTL (High Threshold Level), LTL (Low Threshold Level), or the equivalent level for 3 
dB. It is the accumulated sound averaged during the Run Time. 
 
DOSE                  The accumulated Dose in percent for HTL, LTL, or 3 dB. 
 
TWA                   Time Weighted Average in decibels for HTL and LTL.
 
SEL                     It is the accumulated sound averaged over 8 hours. For
 
3 dB it is the Sound Exposure Level of the accumulated
 
sound averaged over 1 second. 
 
Time Keys 
RUN                    The total time the unit was running and accumulating
 
TIME                   data. If the total time is less than 30 minutes, the co-
 
lon will blink if the key is held down and the time shown is in minutes and seconds. If the 
colon stays on, the time shown is in hours and minutes. 
 
PAUSE                The time the unit was on but not accumulating data.
 
TIME                  The colon blinks as in Run Time 
.
 
U.L.                     The time exposure was above the Upper Limit. The
 
TIME                  colon blinks as in Run Time. 
 
Event Keys 
The event feature accumulates Lavg/Leq, Dose, TWA/SEL for 3 dosimeters, Peak Level, Max 
Level, Run Time, and Start Time for up to 15 different events. Each time a key is pressed in the Event 
mode, the display will show the event number and the data. 
 
EVENT                Enters Events and cycles through the Events with
 
data. The display shows the event number, a "CE" if it is the Current Event or else a "-E" 
 
for a previous Event.
 
As an example: 12CE means 12 is the Current Event, 3-E means 3 was a previous Event. 
Only previous events with data will be displayed. 
 
RUN                    Starts accumulating data for the Current Event. 
 
PAUSE                Stops accumulating data and cycles to the next Event.
 
RESET
 
 
ON/OFF               Exits the Event mode. 
 
Procedure to Accumulate Event Data 
To enter the Event mode, press EVENT. To accumulate data, press RUN. To stop, press 
PAUSE/RESET. Press RUN and PAUSE/RESET again to accumulate another event. Press ON/OFF to exit 
events.
 
The following keys can be used to read data in the current event or in any previous event.
 
CODE                  Cycles through HTL, LTL, and 3 dB. Displays the cur-
 
HL3                     rent time or elapsed time for the current event and
 
the start time for any previous event.