Instruction 3015-4286 
23
 
Maintenance 
Warnings and Cautions 
WARNING: 
 Always disconnect AC power before working inside the monitor.  
 
CAUTION:
  When working inside the monitor, be very careful not to dislodge any 
electrical wiring or pneumatic tubing. The AGMSZ contains sensitive electronic 
components that can be easily damaged. Be careful not to touch or disturb any of these 
components. 
Hydrophobic Filters 
Hydrophobic Filters (Page 8) located in both the gas-sample and purge lines prevent water from entering 
the IR detector. A zone flow fault will occur (fault code <0800>) if the gas-sample filter becomes blocked, 
while a purge flow fault will occur (fault code <1000>) if the purge filter becomes blocked. Replace the 
appropriate filter as required (P/N 07-1650). 
Servicing Air Lines & Termination Filters 
The gas-sample line and the optional purge and exhaust lines should be periodically checked for obvious 
signs of kinks, damage, and contamination. Replace the tubing as required (P/N 304-2742). 
The gas-sample line and purge-line termination filters prevent dust and dirt from entering the monitor. 
Both of these filters should be periodically checked and replaced when there are obvious signs of 
contamination. A zone flow fault will occur (fault code <0800>) if the gas-sample line filter becomes 
blocked, while a purge flow fault will occur (fault code <1000>) if the purge-line filter becomes blocked. To 
service the filter, simply remove it from the line and replace it with a new one (P/N 3015-3420). 
Fuses 
The AGMSZ is protected from electrical damage by two, 1 A, 250 V, type “F” fuses (Page 8). Carefully 
remove the fuses from their fuse clips and visually inspect each fuse for damage. Replace the fuses as 
required (P/N 04-2620). 
Clock Battery 
The Clock Battery (Page 8) maintains the correct date and time when AC power is not applied to the 
monitor. Replace this battery about every 2 years (P/N 204-0020). 
Sample Pump 
The Sample Pump (Page 8) draws the gas sample into the monitor, through the IR detector, and 
discharges the sample out the monitor’s exhaust port. When replacing the pump, remove its inlet and 
outlet tubing, disconnect the AC power wires from the pump itself, and remove pump from monitor. Install 
a new sample pump (P/N 3015-4239).