BiliTx
parents’ manual
3. Setup
 
Setting Up the BiliTx System
1.  Place the appropriate cover on the panel and position the baby 
and panel as described later in this chapter. (See Setup-Wrap 
Around Panel or Flat Neonatal Panel.) 
2.  Place the Illuminator device on a hard, flat surface or using 
the carrying case, hang the unit on the outside of a crib or the 
treatment area, away from any heat source. Make sure it is no 
more than four feet from where the baby will be positioned. 
3.  Insert the metal end of the light panel cable, called the 
ferrule, with the metal post facing up, into the opening on 
the Illuminator unit. Push the ferrule in and rotate it counter-
clockwise to lock into place.  The light will not come on if the 
panel is not inserted in the Illuminator device.
4.  Plug the Illuminator device into an electrical outlet. The power 
button will flash green.
 CAUTION
If the power cord or wires need repair or replacement, do not connect the device.
5.  Press the Therapy On/Off button to turn the Illuminator device 
on and begin phototherapy. The Therapy button will illuminate 
green.
 
 
 
6.  To turn off the device when therapy is complete, press and hold 
the Therapy On/Off button for 3 seconds.
 WARNING
After treatment has begun, the 
baby’s bilirubin level should be 
measured to make sure therapy 
is effective.
 CAUTION
 Do not block any of the air vents 
on the Illuminator device. 
 
Connecting the light panel 
cable to the illuminator 
device
 
Turning on the illuminator 
device