RayTek 54301 用户手册

下载
页码 93
Installation
 
5.4.4 Trigger and Hold Function 
The  FTC3  input  can  be  used  as  external  trigger  in  conjunction  with 
the software trigger mode setting “Trigger” or “Hold”. 
 
External switch: 
- contact relay, 
- transistor, 
- TTL gate, … 
Figure 17: Wiring of FTC3 as External Input 
Trigger: A logical low signal at the input FTC3 will reset the peak or 
valley hold function. As long as the input is kept at logical low level 
the  software  will  transfer  the  actual  object  temperatures  toward  the 
output.  At  the  next  logical  high  level,  the  hold  function  will  be 
restarted. 
Temp 
 
object temperature 
output temperature 
FTC3
Figure 18: FTC3 for Resetting the Peak Hold Function 
Time
28 
MID