6.4.5 Advanced Valley Hold 
This  function  works  similar  to  the  advanced  peak  hold  function, 
except it will search the signal for a local minimum. 
6.4.6 Advanced Peak Hold with Averaging 
The  output  signal  delivered  by  the  advanced  peak  hold  functions 
tends  to  jump  up  and  down.  This  is  due  to  the  fact,  that  only 
maximum points of the otherwise homogenous trace will be shown. 
The  user  may  combine  the  functionality  of  the  peak  hold  function 
with  the  averaging  function  by  choosing  an  average  time,  thus, 
smoothing the output signal for convenient tracing. 
 
output temperature 
object temperature 
without averaging 
Temp 
Time
Figure 29: Advanced Peak Hold with Averaging 
The  advanced  peak  hold  function  with  averaging  is  only  adjustable 
by means of the DataTemp MultiDrop Software. 
6.4.7 Advanced Valley Hold with Averaging 
This function works similar to the advanced peak hold function with 
averaging, except it will search the signal for a local minimum. 