Chapter 
2
File
Menu
300
File – Import – Digitizer – Step Scan (When VIVID 300 is Selected)
8
  Place the calibration chart on the rotat-
ing stage.
Position the chart with the black line facing the 
instrument and place it on the rotating stage 
so that both entire side surfaces are observed 
in the viewfinder.
•  If “Chart” is not displayed in [Current Angle], 
click the left mouse button and select “Chart” 
from the pull-down menu that appears.
M
emo
The calibration chart can be scanned at any angle of the 
rotating stage.
9
  Click the [OK] button.
The calibration chart will be scanned.
Note
Make sure that the acquired range image looks like
the good example given below. If it resembles the bad
example, click the [Scan] button again to re-scan the
calibration chart.
10
 Click the [Convert] button.
The [File-Import-Digitizer-Step Scan-Convert] 
dialog box will appear.
Good example 
Bad example