Chapter 3
Troubleshooting
3-2
ni.com
3.
Depress the circuit breaker to reset it.
4.
Reconnect the AC line cord.
5.
Turn the power switch to the On position.
If the circuit breaker trips again, complete the following steps:
1.
Turn the front panel power switch to the Standby position.
2.
Disconnect the chassis from the AC mains power source.
3.
Remove all modules from the chassis.
4.
Complete the test procedure described in the
section in Chapter 2,
.
5.
If any voltages are outside the acceptable limits, contact National
Instruments.
6.
If all voltages are within the acceptable limits, verify that your
PXI-1006 can meet the power requirements of your CompactPCI or
PXI modules. Overloading the chassis can cause the breaker to trip.
Refer to Appendix A,
7.
The over-current condition that caused the circuit breaker to trip
may be due to a faulty CompactPCI or PXI module. Refer to the
documentation supplied with the modules for troubleshooting your
modules.