Sony SIR-3400H 用户手册

下载
页码 156
 (E) 
8-3
SIR-3400H / SIR-3032i / SIR-3032W
8. Function Setup (Function Screen)
8-2-4. Selecting the Trigger Mode (Trigger Mode)
Factory setting: Single
* For the setting procedure, see “8-1. Setting Procedure (for Function Screen)”.
Settings
Single : During trigger recording, recording starts when the trigger input signal changes from High to
Low, and recording continues until the STOP key is pressed.
Multi : During trigger recording, recording starts when the trigger input signal changes to Low, and
recording stops when it changes to High.
8-2-5. Pre-Trigger Recording On/Off (Pre-Trigger)
Factory setting: Off
* For the setting procedure, see “8-1. Setting Procedure (for Function Screen)”.
<When set to On>
During trigger recording, data before the trigger is recorded. Data length (s) before trigger is automatically
changes depending on to the recording media speed.
<When set to Off>
During trigger recording, and data is recorded immediately after the trigger.
8-2-6. IRIG-B Timecode On/Off (IRIG)
Set this switch to On to calibrate the internal clock by inputting the IRIG-B time code to the AUX analog
channel. When set to Off, the internal clock is not calibrated with the IRIG-B time code.
Factory setting: On