SECTION 2—OPERATION
20
Placing the Patient on the Mattress
1. Press the Mode button or Max Inflate button to turn on the control unit to maximum 
flow. 
NOTE: In this mode, the Max Inflate LED lights up.
2. After the mattress is fully inflated, place the patient on the bed.
3. Ensure that the patient’s feet are towards the foot end of the mattress (the end with the 
hose). 
4. Center the patient on the bed from side‐to‐side and head‐to‐foot.
NOTE: Special positioning may be required with contracted patients to provide comfortable 
positions.
5. After placing the patient, make certain no objects will fall under the patient, such as 
feeding tubes, IVʹs etc.
6. Wait five minutes for the mattress pressure to stabilize.
7. Set the comfort pressure to the desired comfort level. 
8. Verify that the patient has not bottomed out by performing the following steps:
A. Ensure that the patient is lying flat on his/her back in the middle of the mattress.
B. Place four fingers between the air cushions directly underneath the sacral region 
of the patient’s body.
C. Ensure that there is 3 to 4‐finger width clearance between the bottom of the patient 
and the bed frame.
D. Adjust the comfort setting, if needed.
E. Wait five minutes for the mattress pressure to stabilize.
F. Repeat STEPS A‐E until patient has not bottomed out and patient comfort is 
achieved. 
9. If the patient feels that the bed is too soft/hard, press the Soft or Firm button to adjust 
the comfort settings.
10. Use a regular pillow to help support and stabilize the patientʹs head.
Transferring Patient From/To a Gurney
 WARNING
ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney 
before transferring the patient between the bed and the gurney.
1. Engage the wheel locks of the bed. Refer to the owner’s manual provided with the bed.
2. Engage the wheel locks of the gurney. Refer to the owner’s manual provided with the 
gurney.
3. Press the Max Inflate button to achieve maximum mattress pressure.