DIAGNOSTICS 
 
The PG VSI controller on your Alero will provide you with information should a 
malfunction happen with your power chair.  The LED lights on the controller will flash in a 
certain sequence to inform you of a malfunction.  If more than one fault exists, then the 
fault having the highest priority is indicated.  All faults disable the controller and require 
the controller to be turned off then on again if the source of the fault is removed. 
 
 
One Bar 
The batteries need charging or there is a bad connection to the 
batteries.  Check the connection to the batteries.  If the 
connections are good, charge the batteries. 
 
 
Two Bars 
The left side motor has a bad connection.  Check the 
connections to the left side motor.  If the connections are good, 
call your Golden Authorized Dealer for service. 
 
 
Three Bars 
The left side motor has a short circuit to a battery connection.  
Call your Golden Authorized Dealer for service. 
 
 
Four Bars 
The right side motor has a bad connection.  Check the 
connections to the right side motor.  If the connections are 
good, call your Golden Authorized Dealer for service. 
 
 
Five Bars 
The right side motor has a short circuit to a battery connection.  
Call your Golden Authorized Dealer for service. 
 
 
 
Six Bars 
The wheelchair is being prevented from driving by an external 
signal.  The exact cause will depend on the type of wheelchair 
you have; one possibility is the battery charger is connected 
 
 
Seven Bars 
A joystick fault is indicated.  Make sure the joystick is in the 
center position before switching on the control system. 
 
 
Eight Bars 
A control system fault is indicated.  Make sure all connections 
are secure. 
 
 
Nine Bars 
The parking brakes have a bad connection.  Check the parking 
brake and motor connections.  Make sure the control system 
connections are secure. 
 
 
Ten Bars 
An excessive voltage has been applied to the control system.  
This is usually caused by a poor battery connection.  Check 
battery connections. 
 
Note:  If you detect there are problems with the motors, the battery charger, the 
controller or damage to the power wheelchair frame, it is highly recommended 
you contact your local Authorized Golden Dealer.