Theory of Operation
Communications
2
 
2-5 
 
Victoreen Communications Loop – 943-227-15VL 
 
Communication between the detector preamplifier and the UDR or 960 controller may be performed via 
Victoreen loop, a serial asynchronous communications interface.  The communication utilizes Victoreen 
protocol and is intended for use in electrically hostile environments where high noise immunity is required.  
Noise immunity is achieved using differential line driving, and electrically isolating the end of each 
transmit/receive pair with opto-isolators.  Using a ±15 Vdc communications power supply, each 0 to 1 
transition is equivalent to a 30-volt swing in loop voltage.  Refer to the Model 942-200-80-1 
Communications Board instruction manual for further information.   
 
Jumper JP1 controls the communication communications method employed.  JP1, A-B selects 
Victoreen 
loop communications. 
 
RS232C Communications – 943-227-15 
 
Communication between the detector preamplifier and the UDR or 960 controller may be performed via 
RS232C serial asynchronous communications interface.  Jumper JP1 controls the communication 
communications method employed.   
 
JP1, B-C selects RS232C communications. 
 
 
2.5  Preamplifier Switch & Jumper Settings 
 
Switch and jumper settings for the preamplifier are summarized in Table 2-7, 2-8, and 2-9. 
 
Table 2-7.  Preamplifier Serial Interface Board 977-210-10M Switch Settings 
Switch Normal 
Op. 
Switch Function 
SW1 
Open 
Microprocessor Reset switch, momentary contact 
 
 
SW2  
N/A 
Not supplied 
 
 
SW3-1 
ON 
OFF Selects 300 baud  •  1/ON Selects 4800 baud (normal setting
SW3-2 
OFF 
OFF •  (No function is implemented in the 977-210-10M) 
SW3-3 
OFF 
OFF Selects Rate Mode (normal)  • (Selects Maintenance Mode for 
factory PGA and Auto Zero set-up only) 
SW3-4 
OFF 
OFF • (No function is implemented in the 977-210-10M) 
 
 
NOTE 
SW3 is a 4-position DIP switch on the 977-210-10M serial 
interface board.