15 
•  The displays will then show the calibration weight suggested as a 
whole number. For example:    
 
 
 
“LoAd” 
 
“04”   
“KiLoS” 
 
Either put this weight on the platform after the stable symbol is on 
or enter the desired value in whole integer and then press 
[Tare].  
 
•  Place the weight on the platform and press [Tare] again. 
 
•  At all times the scale should be stable before pressing the [Tare] 
key to accept a weight.  The stability indicator will turn on to show 
the value is stable. 
 
•  When calibration is done the display will show “SPAn”  “PASS”. 
 
•  Remove the weight. The scale will return to normal weighing.   
 
•  If  an  error  message  “SPAn”  “FAIL”  is  displayed,  then  repeat  the 
process. 
 
Suggested Calibration weights for CBC Series: 
 
CBC 4 / 
CBC 8a 
CBC 8 / 
CBC 16a 
CBC 16 / 
CBC 35a 
CBC 32 / 
CBC 70a 
CBC 48 / 
CBC 100a 
2 kg / 5 Ib 
 
4 kg / 10 lb 
 
10 kg / 30 lb 
 
20 kg / 50 lb
 
  
30 kg / 100 lb 
 
 
•  After  calibration,  the  scale  should  be  checked  whether  the 
calibration and linearity is correct. If necessary, repeat calibration. 
 
NOTE:  CBCa  scales  will  have  the  lb  or  kg  indicator  on,  to  show  the  unit  of 
the  weight  requested.  If  the  scale  was  in  pounds  before  starting  the 
calibration, the weights requested will be in pound values or if the scale was 
weighing in kilograms then metric weights will be requested. 
  
 
9.0  RS-232 INTERFACE 
 
 
The  CBC  Series  are  supplied  with  a  RS-232  bi-directional  interface.  The 
scale when connected to a printer or computer through the RS-232 interface, 
outputs the weight, unit weight and count.