Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
Section 4: Configuration
3-9000-765 Rev H
May 2015
Using AMS Device Manager to configure the meter
 91
g.
Click 
Service Tools|Variables|All Variables
 tab to view Primary, Secondary, Third 
and Fourth Variable parameter status.
Figure 4-17  AMS Device Manager - Service Tools All Variables  
Gauges display each variable’s status as good or bad. If a status is bad refer to the 
Service Tools Alerts page for recommended actions  to resolve the alert condition. Also 
refer to the Field Device Specification manual (P/N 3-9000-762) for Commands 48 and 
140 details.
Alerts are triggered for Command 48 Additional device status and Command 140 
detailed status information. Alerts are grouped as Failed - Fix Now, Maintenance - Fix 
Soon and Advisory according to the severity level; 1-6. Severity 1 is the highest and 6 is 
the lowest level.