Emerson 3812 User Manual

Page of 138
52
CPU Module labeling and LED indicators
Section 3: Electrical installation
Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
May 2015
3-9000-765 Rev H
Table 3-2  CPU Module labeling and LED functions
CPU Module switch
Function
Switch position indicator or LED
WRITE PROT.
Write-protect mode - with switch in the ON 
position (default setting) protects configuration 
and firmware overwrites. 
To write configuration changes or download 
firmware to the meter change the switch to the 
OFF  position
Switch position
ON -  (default setting) enables write-
protection of the configuration and 
firmware 
OFF - enables writing configuration 
changes or downloading firmware
DHCP
Dynamic Host Protocol Server - enables you to 
communicate with a Daniel meter that is not 
connected to a network. 
When the CPU Module switch is in the ON 
position, the meter is enabled to act as a DHCP 
server for a single DHCP client connected to the 
Ethernet port using a crossover  cable. This should 
be used for peer to peer connections only.
When the connection is made, select to use the 
Meter Name in the meter instead of the Meter 
Directory Name in order to keep all log files and 
configurations separate from each meter.
Switch position
ON -  the meter is enabled to act as a 
DHCP server for a single DHCP client 
OFF - disables the DHCP server
PORT A
PORT A override - RS-232 serves as an override 
during meter commissioning to establish com-
munications and in the event the user cannot 
communicate with the meter due to an inadver-
tent communication configuration change. The 
override period is for two minutes
Supports:
-
auto-detected ASCII (Start bit 1, Data Bit 7, 
Parity Odd/Even, Stop Bit 1) 
-
RTU (Start Bit 1, Data Bit 8, Parity none, Stop 
Bit 1).
-
Modbus protocols
RS-232  Baud rate=19,200
Modbus ID=32
Switch position
ON -  enables RS-232 PORT A override 
OFF - (default setting) disables 
RS-232 PORT A
MEAS
System color indicates metrology mode
Acquisition mode 
 
Measurement  mode
LED status
Red flashing LED
Solid red the Acquisition Module not 
communicating with the CPU Module
Green flashing LED
PWR
3.3V Power Indicator 
Solid Green
LED 4
Not used
LED 5
Not used
RX
RX signal (Port A for RS485 or RS232  
communication) receiving data
Flashing green (when receiving data)
TX
TX signal (Port A for RS485 or RS232  
communication) transmitting data
Flashing green (when transmitting 
data)
LINK
ETH1Link user Ethernet connection
Solid green