Taylor 9867 User Manual

Page of 16
4. Insert the stem of the thermometer at least 2” 
into the container and allow it to stabilize for 5 
minutes. Note: It is important that the tip of 
the thermometer not touch the bottom of 
the container.
5. If the temperature reading is 32°F (0°C), the 
thermometer is accurately calibrated. 
6. If it is necessary to adjust the thermometer, 
press and hold the “CAL” button for 3 seconds. 
The display will show “CAL”.
7. Within 3 seconds, press and hold the “HOLD s
” 
and “°F/°C  t
” buttons at the same time for 5 
seconds. The current temperature blinks. 
8. Press the “HOLD s
” or “°F/°C t
” button to 
increase/decrease the temperature by .5 
degree increments. 
9. When the temperature reading is stable, the 
LCD will stop blinking. Press “CAL” to accept 
and return to temperature mode. The 
thermometer is ready for use.
10.If the temperature reading is not stable, the 
reading blinks “Errc” 3 times and returns to 
temperature mode. Repeat calibration steps.  
Care and Precautions
1. 
Always read instructions thoroughly to ensure 
that you have the full benefit of all the unit's 
features.
2. Sterilize thermometer probe each time before 
use.
3. Use caution while handling the thermometer. 
Take care when moving the probe into position 
to avoid injury or damage. Always wear a heat 
resistant glove when touching the thermometer 
during or just after use. DO NOT TOUCH 
PROBE WITH BARE HANDS!