Princeton 4411-0106 User Manual

Page of 136
32 
PIXIS System Manual 
Version 2.C 
 
6.  When you have finished with focusing and rotational alignment, secure the 
sliding tube in place with the spectrograph setscrews.  
7.  If they have already been installed, loosen the adapter’s two recessed locking 
setscrews (use a 0.050 hex key). If they have not been installed, insert the 
setscrews into the holes at the sides of the adapter and screw them in a couple of 
turns. 
8.  When the camera opening is centered within the sliding tube, you can adjust the 
camera up or down .4” (4.06 mm) by tightening or loosening the vertical 
adjustment setscrew (use a 3/32” hex key).  
9.  After you have completed the adjustment, tighten the recessed locking setscrews 
in the adapter flange to lock the position. 
 
Figure 10.  Adjustable C- to Spectroscopy-Mount Kinetics Adapter 
Mounting the Camera to a Spectrograph 
The camera must be properly mounted to the spectrograph to focus. Additional 
precautions must also be taken to prevent overexposure of the camera. 
The distance to the focal plane from the front of the mechanical assembly depends on the 
specific configuration. Refer to "Focal Distance" on page 93 for more information.  
Spectrograph Adapter 
Refer to Appendix C, page 113 for instructions for a variety of adapters. Other adapters 
may be available. Consult the factory for specific information relating to your needs
.