Princeton 4411-0106 User Manual

Page of 136
Chapter 4 
System Setup 
33 
 
33 
Connecting an External Shutter 
Disconnecting or connecting the shutter cable to the camera while the camera is ON can destroy 
the shutter or the shutter driver in the camera! 
Introduction 
Typically, PIXIS cameras for imaging applications are shipped with an internal shutter. 
The 25 or 45 mm internal shutter (depending on the CCD array size) is housed in the 
main body. Typically, cameras for spectroscopy applications do not incorporate internal 
shutter. However, for cameras without an internal shutter, there is provision for 
connecting an external 25 or 45 mm shutter (supplied by Princeton Instruments) at the 
Shutter connector on the rear of the PIXIS. 
1.  DO NOT connect a Princeton Instruments-supplied external shutter when there is an 
installed internal shutter. Permanent damage to the shutter driver may occur. 
2.  Electromechanical shutters typically have a lifetime of about a million cycles. Avoid 
running the shutter unnecessarily.  
3.  A shutter can become overheated when short, rapidly repeated exposures are used or if 
the shutter is held open for an extended period of time. Newer versions of the 
WinView/WinSpec software and camera hardware monitor the temperature of  45 mm 
internal shutters
 to prevent overheating conditions. The temperature of external 
shutters is NOT monitored!
 
 
External Shutter 
A Princeton Instruments-supplied external shutter may be used with a PIXIS camera that does 
not have an internal shutter. In most cases, the external shutter will be mounted on the 
entrance slit of a spectrograph. The shutter mount used with all Acton Series spectrographs 
requires no disassembly. Mount it to the spectrograph as shown in Figure 11. In the case of 
the IsoPlane SCT-320 spectrograph, the shutter assembly is mounted internally: refer to the 
instructions provided in the IsoPlane manual. 
Shutter Cable Connection 
1.  Verify that the PIXIS camera is turned OFF (i.e., the 
power supply is switched OFF). 
2.  Verify that there is no internal shutter.  
If the camera is equipped with an internal shutter, DO 
NOT USE
 the Shutter connector to drive an external 
(second) shutter. Such a configuration will result in 
under-powering both shutters and may cause damage to 
the system. 
3.  Connect the shutter cable to the LEMO
®
 connector at 
the rear of the camera.   
4.  Power the PIXIS camera ON. 
 
WARNING! 
Figure 11.  Acton Series Spectrograph 
Entrance Slit Mount 
Caution
 
Cautions