Experiment Setup Main tab 
(Acquisition|Experiment Setup…):  
  Exposure Time: 100 ms 
  Accumulations & Number of Images: 1 
Experiment Setup ROI tab 
(Acquisition|Experiment Setup…): Use this 
function to define the region of interest (ROI). 
  Imaging Mode: Select this mode if you are running WinSpec. 
  Clicking on
 
Full loads the full size of the chip into the edit boxes. 
Experiment Setup Timing tab (Acqu
isition|Experiment Setup…): 
  Timing Mode: Free Run 
  Shutter Control: Normal 
  Safe Mode vs. Fast Mode: Fast 
Confirming the Setup 
1.  If you are using WinView/32 and the computer monitor for focusing, select 
Focus
 from the Acquisition menu. Successive images will be sent to the 
monitor as quickly as they are acquired.  
2.  Adjust the lens aperture, intensity scaling, and focus for the best image as 
viewed on the computer monitor. Some imaging tips follow:  
  Begin with the lens blocked off and then set the lens at the smallest 
possible aperture (largest f-stop number). 
  Make sure there is a suitable target in front of the lens. An object with 
text or graphics works best.  
  Adjust the intensity scaling (by clicking the 
5%-95%
 button at the 
bottom left corner of the data window) and adjust the lens aperture until 
a suitable setting is found. Once you’ve determined that the image is 
present, select a lower setting for better contrast. Check the brightest 
regions of the image to determine if the A/D converter is at full-scale. 
16-bit A/D is at full scale when the brightest parts of the image reach an 
intensity of 65535. 
Adjust the aperture to where it is just slightly smaller 
(higher f-stop) than the setting where maximum brightness on any part of 
the image occurs.  
  Set the focus adjustment of the lens for maximum sharpness in the 
viewed image. 
  In the case of a camera with an F-mount, the camera lens adapter itself 
also has a focus adjustment. If necessary, this focus can be changed to 
bring the image into range of the lens focus adjustment. See “F-Mount 
Adapter Focusing Procedure”
, page 110. 
3.  After you have focused the camera, you can stop Focus mode, continue 
Focus mode, begin Acquire mode, or wait for the CCD to reach the 
operating temperature before going to Acquire mode.