  Internal Slit Shutter: A shutter mounted internally has an external shutter 
connector in the sidewall of the spectrograph. Connect a shutter cable from 
the PIXIS Shutter connector to that connector. 
6.  Power ON the camera (i.e., switch the power supply ON).  
Note: With USB 2.0, the camera must be turned on before WinX is opened, and 
WinX must be closed before the camera is turned off. 
7.  Turn on the computer power. 
8.  Start the application software. 
Setting the Camera Parameters 
Note: The following procedure is based on WinSpec/32: you will need to modify it 
if you are using a different application. Basic familiarity with the WinSpec/32 
software is assumed. If this is not the case, you may want to review the software 
manual or have it available while performing this procedure. 
Set the software parameters as follows: 
Environment dialog (Setup|Environment): Check the DMA Buffer size. 
Large arrays (2048x2048, for example) require a buffer size on the order of 
32 Mbytes. If you change the buffer size, you will have to reboot the computer 
for this memory allocation to be activated, and then restart WinSpec. 
Controller|Camera tab (Setup|Hardware): These parameters should be set 
automatically to the proper values for your system.  
  Controller type: This information is read from the camera. 
  Camera type: This information is read from the camera.  
  Shutter type: Remote (entrance slit shutter). You would select None if the 
camera had no shutter and was not controlling an entrance slit shutter. 
  Readout mode: Available modes are read from the camera. Select Full 
frame. 
Detector Temperature (Setup|Detector Temperature…):  The default 
temperature setting is read from the camera. When the array temperature 
reaches the set temperature, the 
Detector Temperature
 dialog will report 
that the temperature is 
LOCKED
. Note that some overshoot may occur. This 
could cause temperature lock to be briefly lost and then quickly re-
established. If you are reading the actual temperature reported by the 
application software, there may be a small difference between the set and 
reported temperature when lock is established. This is normal and does not 
indicate a system malfunction. Once lock is established, the temperature will 
be stable to within ±0.05°C. 
Note:
 
The Detector Temperature dialog will not display temperature 
information while you are acquiring data. 
Cleans and Skips tab (Setup|Hardware): Click on 
Load Default Values
 
and click on 
Yes