Princeton 4411-0106 User Manual

Page of 136
Chapter 5 
Operation 
41 
 
41 
Experiment Setup Main tab (Acquisi
tion|Experiment Setup…):  
  Exposure Time: 100 ms 
  Accumulations & Number of Images: 1 
Experiment Setup ROI tab 
(Acquisition|Experiment Setup…): Use this 
function to define the region of interest (ROI). 
  Spectroscopy Mode: Selected 
  Clicking on
 
Full loads the full size of the chip into the edit boxes. 
Experiment Setup Timing tab 
(Acquisition|Experiment Setup…): 
  Timing Mode: Free Run 
  Shutter Control: Normal 
  Safe Mode vs. Fast Mode: Fast 
Setting the Spectrograph Parameters 
1.  Define Spectrograph dialog (Spectrograph|Define): Click on 
Install/Remove Spectrograph
, highlight the appropriate spectrograph name in the 
Supported Spectrographs
 list (for example, Acton SP-300i for an Acton 
SP-2300i or Acton SCT320 for an IsoPlane), and click on 
Install Selected 
Spectrograph
2.  Move Spectrograph dialog: Choose the grating to be move, and then set it to 
500 nm if using a mercury lamp or to 0.0 nm if using a broadband source. 
Confirming the Setup 
1.  Turn on the light source at the spectrograph entrance slit. 
2.  In WinSpec, select 
Focus
 (on the Acquisition menu or on the 
Experiment Setup
 
dialog) to begin data accumulation. Depending on the display settings, you should 
see either a spectral band (image) or a graph. Background noise will decrease as the 
camera cools to its default temperature. 
3.  Turn off the light source. The data displayed should change to a background noise 
pattern or low intensity graph. If this occurs, you have confirmed that light entering 
the spectrograph is being seen by the camera. Skip Steps 4-5 and continue to the 
Focusing” topic on page 42. 
4.  If there is no difference between the data displayed when the light source is on or off: 
a.  Verify that the light source has power and is turned on. 
b.  Verify that the entrance slit is open at least 10 µm.  
c.  Check the 
Exposure Time
 (
Experiment Setup Timing
 tab
)
.  
d.  Confirm that 
Shutter Control
 is set to Normal (
Experiment Setup Timing
 tab
)
.  
e.  Check the shutter cable connections. 
f.  Verify shutter operation. You should hear the shutter open and close while 
running in 
Focus
 
mode.  
  If you hear a shutter operating and you have done Steps a-e, turn the light source 
on, wait a minute and then turn the light off while you view the data display. If