For signal levels low enough to be readout-noise limited, longer exposure times, and 
therefore longer signal accumulation in the CCD, will improve the S/N ratio 
approximately linearly with the length of exposure time. There is, however, a maximum 
time limit for on-chip accumulation, determined by either the saturation of the CCD by 
the signal or the loss of dynamic range due to the buildup of dark charge in the pixels. 
Readout 
Introduction 
After the exposure time has 
elapsed, the charge 
accumulated in the array pixels 
needs to be read out of the 
array, converted from electrons 
to digital format, and 
transmitted to the application 
software where it can be 
displayed and/or stored. 
Readout begins by moving 
charge from the CCD image 
area to the shift register. The 
charge in the shift register 
pixels, which typically have 
twice the capacity of the image  
 
Figure 24.  Array Terms for a CCD with a 
 Dual Output Amplifier 
pixels, is then shifted into the output node and then to the output amplifier where the 
electrons are grouped as electrons/count. This result leaves the CCD and goes to the 
preamplifier where gain is applied.  
WinX and LightField allow you to specify the type of readout (full frame or binned), the 
output amplifier, and the gain (the number of electrons required to generate an ADU).