Chapter 
7
 
Troubleshooting 
87 
 
87 
Cooling Troubleshooting 
Temperature Lock Cannot be Achieved or Maintained.  
Possible causes for not being able to achieve or maintain lock could include: 
  Ambient temperature greater than +23°C. This condition affects TE-cooled 
cameras. If ambient is greater than +23°C, you will need to cool the camera 
environment or raise the set temperature.   
  Airflow through the camera and/or circulator is obstructed. The camera needs to 
have approximately two (2) inches (50 mm) clearance around the vented covers. 
If there is an enclosure involved, the enclosure needs to have unrestricted flow to 
an open environment. The camera vents its heat out the vents near the nose. The 
air intake is near the rear of the camera. 
  A hose is kinked. Unkink the hose. 
  Coolant level is low. Add coolant. See “CoolCUBE
II
: Low Coolant (Air in Hoses)” 
on page 86. 
  There may be air in the hoses. Add coolant. See “CoolCUBE
II
: Low Coolant 
(Air in Hoses)” on page 86. 
  Circulator pump is not working. If you do not hear the pump running when the 
CoolCUBE
II
 is powered on, turn off the circulator and contact Customer Support. 
  The circulator is higher than the camera. Reposition the circulator so that it is 6 
inches (150 mm) or more below the camera. The vertical distance should not exceed 
10 feet (3 m). Typically, the camera is at table height and the circulator is on the 
floor. 
  The camera vacuum has deteriorated and needs to be refreshed. Contact 
Customer Support. 
  The target array temperature is not appropriate for your particular camera and CCD 
array.   
  For a TE-cooled camera, the camera's internal temperature may be too high, such as 
might occur if the operating environment is particularly warm or if you are 
attempting to operate at a temperature colder than the specified limit. TE-cooled 
cameras are equipped with a thermal-protection switch that shuts the cooler circuits 
down if the internal temperature exceeds a preset limit. Typically, camera operation 
is restored automatically in about ten minutes. Although the thermo-protection 
switch will protect the camera, you are nevertheless advised to power down and 
correct the operating conditions that caused the thermal-overload to occur. 
Camera loses Temperature Lock 
The internal temperature of the camera is too high. This might occur if the operating 
environment is particularly warm or if you are trying to operate at a temperature colder than 
the specified limit. If this happens, an internal thermal overload switch will disable the 
cooler circuits to protect them. Typically, camera operation is restored in about ten minutes. 
Although the thermal overload switch will protect the camera, users are advised to power 
down and correct the operating conditions that caused the thermal overload to occur.