Emerson Process Management 3422 User Manual

Page of 398
 Reference, Installation, and Operations Manual
Troubleshooting comm., mech., and elec.
3-9000-743 Rev S
June 2013
MARK III mechanical/electrical troubleshooting
 315
D.2 MARK III mechanical/electrical troubleshooting
This section is meant to assist site maintenance and operations personnel trained in the 
operation of the ultrasonic flow meter and knowledgeable in basic mechanical and electronic/
electrical troubleshooting techniques, using lap top computers as well as digital volt/ohm 
meters. Great care should be taken not to “short out” a given electronic/electrical circuit when 
Troubleshooting.
Table D-1  Mark III Mechanical/Electrical Troubleshooting
Problem
Solution(s)
no power to the unit
Check for correct voltage (AC or DC) to the input of the Field Connection Board. 
(
, System Wiring Diagram).
Check the main power source for blown fuse or tripped circuit breaker. Reference 
your “as built” installation drawings for your location. 
Check the fuses on the Field Connection Board. Reference fuse F1 and F2 locations.
cannot communicate with Daniel 
MeterLink program
Ensure that the meter is properly powered. 
Ensure that the computer cable is properly connected to the field connection board 
(
) for the connector and check your interface pins (RS-485 or RS-
232).
Verify that the communication parameters of the Daniel MeterLink program are set 
according to jumpers on the meter CPU board. 
 of this manual for 
instructions on configuring communications.
Check switch S-1 on the field connection board and ensure that it is in the correct 
position 
(RS-485 or RS-232 communication). Note that S1 is located at the top left side of 
the board. 
one or more of the chords is not 
indicating a reading (reporting 
zeros)
Check for loose connections at the cable connectors. (
).
Check the resistance of the transducers (should be approximately 1-2 
Ω
).
Problem also may be caused by a bad Acquisition board or interconnect cable. 
 drawing for more information
Check system status in the Daniel MeterLink program for any flagged errors. 
 for system status information.
Check the CPU board. 
wave form contains an excessive 
amount of noise
Increase the 
StackSize
 till noise level decreases (settings can be 1 (none), 2, 4, 8, or 
16). 
 for instructions on changing register values. If increasing the 
StackSize
 is not successful, try turning on the filter or consult with Daniel Customer 
Service if you are unsure of how stacking a signal can effect the meter's operation. Refer 
to Technical Support under the Help menu of Daniel MeterLink for contact information.
connected communication line to 
the flow computer but no signal is 
received
Check for loose connections at the flow meter and the flow computer (
).
Check the switch settings on the CPU board, Field Connection board and the Option 
board. Make sure the switches are in the correct position (
 through 
 for RS-485 or RS-232 communication).