Emerson Process Management 3422 User Manual
316
MARK III mechanical/electrical troubleshooting
Troubleshooting comm., mech., and elec.
Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
communicating with meter but all
chords display failures
chords display failures
•
Verify that the resistance of transducers is within Specification (1-2
Ω)
.
•
Check the Acquisition board (
).
•
Check the interconnect cables between the base assembly and the head assembly
(
(
chord is not indicating
•
Check the resistance of the failed transducer.
•
If Chord A is not indicating, change the transducer cables from Chord D to chord A.
If Chord D then fails, the transducers are bad on Chord A (
If Chord D then fails, the transducers are bad on Chord A (
).
•
The same test procedure can be accomplished by swapping Chords B and C if a
chord fails in either chord (
chord fails in either chord (
Note that the outside chord cables cannot be exchanged with inner
chord cables.
chord cables.
Table D-1 Mark III Mechanical/Electrical Troubleshooting
Problem
Solution(s)