Emerson Series 3000 MVD User Manual

Page of 348
Configuration and Use Manual
215
Diagnostics and Troubleshooting
Dia
g
nostics and 
T
rou
b
lesh
oo
tin
g
Sp
eci
ficat
ion
s
 – Mo
del 
3350
/3
700
Specific
a
tions
 –
 Model 
330
0
/3
500
Measure
ment P
e
rf
or
mance
22.7.2
Sensor alarms
Table 22-5 lists all alarms in the Sensor category, with descriptions, suggested user actions, and 
related information.
A130
Printer Out of 
Paper
(Custody transfer 
only) The ticket 
printer is out of 
paper.
Ignore
Yes
No
Add paper to the printer.
A134
PPI Memory 
Error
Memory or 
checksum error 
in transmitter 
display.
Info
Yes
No
Cycle power to the 
flowmeter. If the problem 
persists, contact Micro 
Motion customer service.
A135
PPI Comm Error
Internal 
communication 
error with 
transmitter 
display.
Info
Yes
No
Cycle power to the 
flowmeter. If the problem 
persists, contact Micro 
Motion customer service.
(1) Applies only to systems with the standard core processor.
(2) Applies only to systems with the enhanced core processor.
(3) Applies only to systems with transmitter software earlier than rev7.0.
Table 22-5
Sensor alarms
Alarm 
number
Maintenance 
menu listing
Description
Alarm severity
Affected 
by fault 
timeout
User actions
Default
User 
config?
A003
Sensor Failure
No tube vibration 
detected at the 
sensor.
Fault
Yes
Yes
Check the test points. See 
Section 22.18.
Check the sensor coils. See 
Section 22.20.
Check wiring to sensor. See 
Section 22.11.2.
Check for slug flow. See 
Section 22.7.3.
Check sensor tubes.
Table 22-4
Electronics alarms continued
Alarm 
number
Maintenance 
menu listing
Description
Alarm severity
Affected 
by fault 
timeout
User actions
Default
User 
config?