Emerson Series 3000 MVD User Manual
216
Micro Motion
®
Series 3000 MVD Transmitters and Controllers
Diagnostics and Troubleshooting
22.7.3
Process alarms
Table 22-6 lists all alarms in the Process category, with descriptions, suggested user actions, and
related information. More information on specific process alarms is provided later in this section.
related information. More information on specific process alarms is provided later in this section.
A004
Temp.
Out-of-Range
Out-of-Range
The measured
temperature is
outside the
sensor limits.
temperature is
outside the
sensor limits.
Fault
No
Yes
Check wiring to sensor. See
Section 22.11.2.
Section 22.11.2.
Verify sensor configuration.
See Section 7.3.3.
See Section 7.3.3.
Check the test points. See
Section 22.18.
Section 22.18.
Check the sensor coils and
RTD. See Section 22.20.
RTD. See Section 22.20.
Verify that process
temperature is within range
of sensor and transmitter.
temperature is within range
of sensor and transmitter.
Contact Micro Motion
customer service.
customer service.
A016
Sensor RTD Error Sensor RTD has
failed.
Fault
Yes
Yes
Check wiring to sensor. See
Section 22.11.2.
Section 22.11.2.
Check the test points. See
Section 22.18.
Section 22.18.
Check the sensor coils. See
Section 22.20.
Section 22.20.
Contact Micro Motion
customer service.
customer service.
A017
Meter RTD Error
Meter RTD has
failed.
failed.
Fault
Yes
Yes
Check wiring to sensor. See
Section 22.11.2.
Section 22.11.2.
Contact Micro Motion
customer service.
customer service.
Table 22-5
Sensor alarms continued
Alarm
number
number
Maintenance
menu listing
menu listing
Description
Alarm severity
Affected
by fault
timeout
by fault
timeout
User actions
Default
User
config?
config?