Diagnostics and Troubleshooting
22.7.3
Process alarms
Table 22-6 lists all alarms in the Process category, with descriptions, suggested user actions, and 
related information. More information on specific process alarms is provided later in this section.
A004
Temp. 
Out-of-Range
The measured 
temperature is 
outside the 
sensor limits.
Fault
No
Yes
Check wiring to sensor. See 
Section 22.11.2.
Verify sensor configuration. 
See Section 7.3.3.
Check the test points. See 
Section 22.18.
Check the sensor coils and 
RTD. See Section 22.20.
Verify that process 
temperature is within range 
of sensor and transmitter.
Contact Micro Motion 
customer service.
A016
Sensor RTD Error Sensor RTD has 
failed.
Fault
Yes
Yes
Check wiring to sensor. See 
Section 22.11.2.
Make sure the appropriate 
sensor type is configured. 
See Section 7.3.3.
Check the test points. See 
Section 22.18.
Check the sensor coils. See 
Section 22.20.
Contact Micro Motion 
customer service.
A017
Meter RTD Error
Meter RTD has 
failed.
Fault
Yes
Yes
Check wiring to sensor. See 
Section 22.11.2.
Make sure the appropriate 
sensor type is configured. 
See Section 7.3.3.
Contact Micro Motion 
customer service.
Table 22-5
Sensor alarms continued
Alarm 
number
Maintenance 
menu listing
Description
Alarm severity
Affected 
by fault 
timeout
User actions
Default
User 
config?