7. SERIAL COMMUNICATIONS  
The 460L comes equipped with a powerful digital Serial Communications Port that can be used for 
Data Acquisition and for changing the monitor’s configuration.  This port can be configured for 
either RS-232 or RS-485 (half-duplex) operation.  See Section 3.7.1 for details on configuring the 
port and connecting it to a computer or data acquisition system. 
7.1. Serial Port Command Syntax 
All characters sent and received are standard ASCII characters and all numbers are decimal 
numbers converted to ASCII text. 
All commands are sent using the following syntax: 
<address><command>:<data1>,<data2>#<checksum (optional)><CR> 
Where: 
  address    is the monitor address (default =1) 
  command   is the command string being sent 
  : (colon)   is the data separator and is only included if data is being sent as part of the 
command (See Command Details below to see if a command requires data or 
not) 
  data1 
is the first data parameter, if required. 
  data2  
is the second data parameter, if required. 
  #  
is the Checksum separator, sent only if optional checksum is included 
The checksum is an ASCII checksum of all characters up to the # character.  
The checksum is optional.  Commands sent without the checksum (and 
checksum separator,) are also valid. 
  CR  
is a carriage return, ASCII 13. 
Examples 
Valid Commands with no data: 
  Checksum Included:  1ALMACK#474<CR> 
  No Checksum:   1ALMACK<CR> 
Valid Commands with data: 
  Checksum Included:  1VSET:1,20#620<CR> 
  No Checksum:  1VSET:1,20<CR>