7. SERIAL COMMUNICATIONS
The 460L comes equipped with a powerful digital Serial Communications Port that can be used for
Data Acquisition and for changing the monitor’s configuration. This port can be configured for
Data Acquisition and for changing the monitor’s configuration. This port can be configured for
either RS-232 or RS-485 (half-duplex) operation. See Section 3.7.1 for details on configuring the
port and connecting it to a computer or data acquisition system.
port and connecting it to a computer or data acquisition system.
7.1. Serial Port Command Syntax
All characters sent and received are standard ASCII characters and all numbers are decimal
numbers converted to ASCII text.
numbers converted to ASCII text.
All commands are sent using the following syntax:
<address><command>:<data1>,<data2>#<checksum (optional)><CR>
Where:
address is the monitor address (default =1)
command is the command string being sent
: (colon) is the data separator and is only included if data is being sent as part of the
command (See Command Details below to see if a command requires data or
not)
data1
is the first data parameter, if required.
data2
is the second data parameter, if required.
#
is the Checksum separator, sent only if optional checksum is included
The checksum is an ASCII checksum of all characters up to the # character.
The checksum is an ASCII checksum of all characters up to the # character.
The checksum is optional. Commands sent without the checksum (and
checksum separator,) are also valid.
checksum separator,) are also valid.
CR
is a carriage return, ASCII 13.
Examples
Valid Commands with no data:
Checksum Included: 1ALMACK#474<CR>
No Checksum: 1ALMACK<CR>
Valid Commands with data:
Checksum Included: 1VSET:1,20#620<CR>
No Checksum: 1VSET:1,20<CR>