Invacare IRC 1720 User Manual

Page of 124
SECTION 3—BATTERIES
25
Using the Portable Nebulizer Battery
1. Align the rubber feet of the compressor with the 
indentations on the battery pack.
2. While holding the battery pack and compressor, push in 
the two tabs on the side of the battery pack.
3. Gently pull the compressor and battery pack in opposite 
directions to ensure the tabs have seated properly. 
NOTE: The GREEN light on the battery will not illuminate while 
the nebulizer is being powered. The RED light will turn on when the 
battery power level is low. When the RED light is on, immediately 
find an alternative power source.
4. To remove the battery pack, press in the two tabs on the 
side of the battery pack and pull the two pieces apart.
Charging the Battery
The battery can be charged with either the power cable (AC) 
or the auto adapter cable (DC). Plug the cable into the 
appropriate power source. The ORANGE battery light will 
turn on indicating that the charging cycle has begun. The 
battery will reach full charge after at least 16 hours.