mA Output (P200 to P219)
When an mA Output Parameter is accessed, the mA symbol is displayed in the Index Type 
field and the mA output number (corresponding to the SITRANS LU01 terminals) is 
displayed in the Index field.
P200 mA Output Range 
Determines the mA output range.
If either 1 or 2 is selected, the mA output is directly proportional to the mA Function. If 
either 3 or 4 is selected, then the output is inversely proportional.
P201 mA Output Function 
Alters the mA output/measurement relationship.
P203 mA Output Value / Transducer 
Displays current mA output value.
This displays as an Auxiliary Reading when 
 key is pressed in the RUN mode and 
does not include adjustments made using Trim features (P214 / P215).
Primary Index
mA output
Values
0
off
1
0 to 20 mA
20 mA = high level (unless P001 = 2 
or 3, then 20 mA = low)
2
*
4 to 20 mA
3
20 to 0 mA
20 mA = low level (unless P001 = 2 
or 3, then 20 mA = high)
4
20 to 4 mA
Related
P911 mA Output Value
Primary Index
mA output
Values
value
mA function
Operation (P001)
0
OFF
1
level
level
2
space
space
3
distance
distance
4
volume
Related
P202 mA Output Allocation
P911  mA  Output  Value
Altered By
P001  Operation
Values
Range: 0.000 to 22.00 (view only)
Note: 
This parameter is applicable only if any mA output has the transducer 
Index Number as its input source (see P201).